41-2 Unpacking and Inspection CAUTION Handle this device carefully at all times. Strong shock to the device may cause trouble during operation.
51-3 Specifications Table 1 Specifications Performance specifications General Power supply 120VAC 60 Hz (For USA area) 230VAC 50 Hz (For Europe a
62. PART NAMES 2-1 Main Unit Fig.2 TM-2655/TM-2655P whole view Note The printer unit is available only for the TM-265
72-2 Rear Panel Fig.3 TM-2655/TM-2655P rear panel (The illustration above is not to scale.) 2-3 Symbols The symbols used with the TM-2655
8What is An Irregular Heartbeat? The TM-2655P provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregu
93.INSTALLATION Follow the procedure below to install the TM-2655/TM-2655P. 1. To ensure that the TM-2655/TM-2655P works properly, install the devic
104. OPERATION 1. Turn on the POWER switch located on the rear panel. When the POWER switch is turned on, all of the display symbols will appear f
118. When the measurement is complete, the air is automatically released from the cuff and the cuff will become loose. 9. The measurement results
125.SETTING THE CLOCK 5-1 Clock Adjustment Mode The clock is set in the clock adjustment mode. The clock adjustment mode display and switches used ar
138. Press the switch to display “14”. 9. Press the SELECT switch to make the minutes section of the TIME display flash. 10. Press the switch
© 2005 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any l
146. INSTALLING THE PRINTER PAPER (This chapter is applicable only to the TM-2655P.) With the power turned on, install the printer paper as follow
157. SETTING THE FUNCTIONS 7-1 Function Setting Mode The TM-2655/TM-2655P is provided with various functions, which allow the user to set the devic
167-2 Description of the Functions Table 5 Functions No. Function SYSTOLIC display DIASTOLIC display Description F01 f01 off Not available. off No
177-3 Printing Samples High-speed printing Graph printing 4-line printing Table printing
188. COMMUNICATION SPECIFICATIONS The TM-2655/TM-2655P is equipped with two RS-232C channels. Various settings for each channel are available in the
198-2 Channel 2: D-sub 9-pin CAUTION The personal computer and medical equipment connected to the device are not allowed to be in the patien
209. MAINTENANCE 9-1 Replacing the Arm Cuff Cover Replace the arm cuff cover as follows: 1. Locate the cover at the bottom of the arm insertion sect
219-2 Replacing the Fuses Replace the fuses as follows: 1. Remove the caps of both fuse holders located on the rear panel. 2. Replace both fuses
229-4 Error Codes Table 9 Error codes Error code Description pe The printer paper has run out. Install a new roll of printer paper. HU The printer
239-7 Options and Consumables Options and consumables available for the TM-2655/TM-2655P are shown below. Order them from the nearest dealer. Use th
iWARNING DEFINITIONS The warnings described in this manual have the following meanings: WARNING Important information to alert you to a situati
2410. TROUBLESHOOTING If the TM-2655/TM-2655P does not function properly or an error code appears, try the following corrective actions. Table 10
2511.EXTERNAL DIMENSIONS Fig.20 External dimensions Note The illustration above indicates the TM-2655P with the printer unit
26MEMO
28MEMO
iiCOMPLIANCE Compliance with the European Directive 93/42/EEC for Medical Products This device conforms to the following requirements: European Direc
iiiSAFETY PRECAUTIONS WARNING Grounding To avoid electrical shock, connect the power cable to a medical electrical outlet having a ground termin
ivGENERAL PRECAUTIONS Obey the following precautions for safe and correct usage. Precautions unique to the TM-2655/TM-2655P appear on the relevant
1CONTENTS WARNING DEFINITIONS...ⅰ COMPLIANCE...
29. MAINTENANCE.......20 9-1 Replacing the Ar
31.INTRODUCTION 1-1 Features The TM-2655/TM-2655P is a blood pressure monitor that measures systolic and diastolic blood pressure and pulse. The TM-2