Weighing Indicator for Check Weighing Scales 1WMPD4002313
8 5. NAMES AND FUNCTIONS No. NAMES FUNCTIONS / OPERATION WEIGHT DATA DISPLAY Indicates the weight data.
9 No. NAMES FUNCTIONS / OPERATION COMPARISON INDICATORS Indicates the results of the weight comparison LO (RED), OK (GREEN) and HI (YELLOW).
10 6. FUNCTION PARAMETER SETTINGS To design a platform scale, set the capacity, weighing unit and other functions according to the specification
11 6-2. Function list Item Function number Description C01- 0 CAL switch enabled, 0 key enabled C01- 1 CAL switch disabled, 0 key
12 7. CALIBRATION This function adjusts the scale (indicator) for accurate weighing. Calibration must be done when the FS-D is initially connect
13 If you need ZERO calibration but do not need SPAN calibration, turn the power off to exit from the calibration procedure. 7. To calibrate wi
14 8. AUTO-TARE FUNCTION The FS-D has an auto-tare function to be used with the comparator function enabled. If the weight value is within the OK
15 9. ADDITIONAL INFORMATION This section describes the additional functions and corrections to the instruction manual of FS-i series (WM+PD40013
16 The frame ground and the shield are connected internally and there is no difference between them in use. Data format Example of data Out of
17 Command H I , + 0 0 1 0 0 CRLF Assume 2 decimal place number. Reply H I , + 0 0 1 0 0 CRLF 1.00% will be set as LO limit %. Command L O
This Manual and Marks All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard
18 10. ERROR MESSAGES Overload error Warning to indicate that an object beyond the weighing capacity has been placed on the weighing pan. Remove
19 11. SPECIFICATIONS 11-1. Capacity and minimum display C02- setting 0* 1 2 3 4 5 6 7 Capacity 6 15 30 60 150 300 600 1500 0.002* 0.005 0.01 0
20 11-3. Options OP-02 (HC-02i) SLA Battery (Yuasa Battery NP4-6 recommended.) OP-03 (FS-03i) RS-232C + Comparator Relay Output (See note.) OP-04
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, IN
1 Contents 1. COMPLIANCE ... 2 2. INTRODUCTION ...
2 1. COMPLIANCE Compliance with FCC rules Please note that this equipment generates, uses and can radiate radio frequency energy. This equipme
3 2. INTRODUCTION This manual describes how this product works and how to get the most out of it in terms of performance. The FS-D weighing indi
4 3. UNPACKING Unpack the indicator carefully and keep the packing material if you are likely to transport the indicator again in the future. W
5 4. INSTALLATION and PRECAUTIONS 4-1. Precautions for Installing the Indicator Ground the indicator so that the user will not be subjected an
6 Setscrew BSetscrew A 4. Solder the wires to the pin contacts of the connector. A 5 m or less load cell cable is recommended as the FS-D uses
7 4-4. Setting Up the Indicator Connect the main power cord to an outlet that has an earth ground terminal. You may use the ear