SV-A series
-Stand Type-
SV-H series
-Portable Type-
Vibro Viscometer
INSTRUCTION MANUAL
1WMPD4001459A
Page view 0
1 2 3 4 5 6 ... 101 102

Summary of Contents

Page 1 - SV-H series

SV-A series -Stand Type- SV-H series -Portable Type- Vibro Viscometer INSTRUCTION MANUAL 1WMPD4001459A

Page 2

8Windows communication toolsWinCT-Viscosity(CD-ROM) (1 pc)Sample cup (Lid attached) (Polycarbonate, Capacity: 2 mL, 10 pcs)Water jacket (1 pc)Connec

Page 3 - CONTENTS

98120741321122912381325717013030314718. EXTERNAL DIMENSIONS SV-1H / SV-10H / SV-100H Display unit (For all models) Sensor unit (Handle*

Page 4

99 SV-1A / SV-10A / SV-100A Stand unit* * The SV-H series external dimensions are as follows when the AX-SV-51 (stand set) is used.

Page 5 - 1. INTRODUCTION

100 Detailed View of the Sensor Unit SV-1A/SV-1H SV-10A/SV-10H SV-100A/SV-100H

Page 6

9 Accessories for SV-10A / SV-100A . Note Please confirm that the AC adapter type is correct for your local

Page 7 - 1-2 Features

10Unpacking Sample cupstand (1 pc)Glass samplecup[2ml] (10 pcs)Sampl

Page 8

11Assembling Note: When assembling, handle the sensor plates with care. When removing or tightening

Page 9 - 2. UNPACKING THE VISCOMETER

12Connection cablefrom the display unit(Connect as shown)Main unitGuide channelSerial numberLeverCapHandle the sensorplates with care. 5 Raise the l

Page 10 - Accessories for SV-1A

132-2 Part names for the SV-H series (Portable type) Accessories Note Please confirm that the AC adapter type i

Page 11 - Polycarbonate

14Unpacking / Assembling Note: When assembling, handle the sensor plates with care. 1 Unpack the parts

Page 12 - SV-10A / SV-100A

15 4 Confirm that the stand’s mounting holes and the display’s mounting holes match up. Secure both sides of the display with the knobs. 5

Page 13 - Install from forward

163. CHANGING TYPE 3-1 Changing from SV-A (Stand type) to SV-H (Portable type) In the case that you have purchased an SV-A series (stand type), an

Page 14 - Display unit

17ScrewsBaseSupporting post 6 Press the START or HOLD key to display Htype . 7 Press the PRINT key to store the setting. After the viscometer

Page 15

© 2008 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any l

Page 16 - Sensor plate

18 2 Remove the attachment screw from the base. Install the X-Y-Z stage along the guide ribs from forward of the base. Secure the X-Y-Z stage using

Page 17

19 5 Insert the AC adapter plug into the AC adapter jack located on the rear side of the display unit. Insert the other end of the AC adapter plug i

Page 18 - 3. CHANGING TYPE

204. DISPLAY AND KEYS 4-1 Display Name Description Standby mode Displays [- - - - -]. Measurement mode Displays the viscosity value in

Page 19 - From the previous page

214-2 Keys (For all models) Key Description ON:OFF Power Turns the power on and off. When the power is turned on, the viscometer enters the

Page 20

224-3 Displaying the Viscosity Values The viscosity values are displayed as below, depending on the unit selected and the viscosity range. The corre

Page 21

234-3-2 SV-10A/SV-10H Use the MODE key to switch between mPa⋅s (Millipascal second) and Pa⋅s (Pascal second), or between cP (Centipoise) and P (Poise

Page 22 - 4. DISPLAY AND KEYS

245. PRECAUTIONS To get the optimum performance from the viscometer and acquire accurate measurement data, note the following: 5-1 General Precautio

Page 23 - 4-2 Keys (For all models)

25121Protector Rotationalaxis 5-3 During Use (For all models)  The SV-A/SV-H series viscometer, designed for very accurate measurement, is sensitiv

Page 24 - 4-3-1 SV-1A/SV-1H

265-4 After Use (For all models)  Remove any residual sample material from the sensor plates, temperature sensor and protector using alcohol. Using

Page 25 - 4-3-3 SV-100A/SV-100H

275-5 Measuring the Absolute Value of Viscosity (For all models) The SV-A/SV-H Series Sine-wave Vibro Viscometer, as a measuring principle, detects

Page 26 - 5. PRECAUTIONS

1CONTENTS 1. INTRODUCTION ...

Page 27

285-5-2 At Calibration When calibrating, enter the product of the absolute viscosity value and the density of the standard viscosity fluid used for c

Page 28 - Temperature

29 6. MEASUREMENT 6-1 Preparing the Sample (For the SV-A series) * Vibration or shock during transportation may cause a display value error. Before

Page 29 - 5-5-1 At Measurement

307 Turn the knob (Height direction) so as to adjust the sample surface to the center of the narrow part of the sensor plates. Confirm that the samp

Page 30 - 5-5-2 At Calibration

31NO6 Confirm that the positioning of the sensor plates is in center of the sample cup by adjusting the X-Y-Z stage. 7 Turn the knob (H

Page 31 - 6. MEASUREMENT

326-2 Basic Measurement Procedure (For the SV-A series) 1 With the power turned off, press the ON:OFF key. When the display is in the standby m

Page 32 - Approx. 25cm

336-3 Basic Measurement Procedure (For the SV-H series) 1 Pour the sample into the beaker (500ml) etc. 2 With the power turned off, press the O

Page 33

34 - With SV-10H/100H, use the surface locator plate as a guide. The surface locator plate has been secured in position so that the tip of the surfac

Page 34 - Below is an exam

35 5 After measurement, the buzzer sounds and the displayed value blinks, then holds. Note When the values do not stabilize because of hand moveme

Page 35

366-4 Changing Units (For all models) The units of viscosity available are: mPa⋅s (millipascal second), Pa⋅s (Pascal second), cP (Centipoise), P

Page 36

377. USING THE WATER JACKET 7-1. Introduction The water jacket, is used with the SV-A Series Vibro Viscometer, to maintain the sample temperature co

Page 37 - (Data hold

28. VISCOSITY CALIBRATION (FOR ALL MODELS) ...42 8-1 Notes on Viscosity Calibrati

Page 38 - SV-100A/SV-100H

387-2. Installation With the SV-1A, as the viscometer, the sample cup holder and the sample cup (capacity: 2 mL) provided or the glass sample cup (ca

Page 39 - 7. USING THE WATER JACKET

39 4 Model: SV-1A Insert the sample cup into the sample cup holder, ant then into the water jacket indicated as 3. When the specific gravity of the

Page 40 - 7-2. Installation

407-3. How to Use 1 Attach the water jacket on the table along the guides. Make sure that the water jacket is slid furthest along the guides until s

Page 41

417-4. Measuring the Absolute Value of Viscosity Using the Small Sample Cup The SV-10A/100A has been calibrated using the accessory sample cup (45 mL

Page 42 - 7-3. How to Use

428. VISCOSITY CALIBRATION (FOR ALL MODELS)  With the SV-H series, when calibration is needed, use the stand set (AX-SV-51, sold separately)  Vis

Page 43 - 7-6. Specifications

438-1 Notes on Viscosity Calibration (For all models)  Pay close attention to the liquid temperature at calibration. Be sure to enter the temperatu

Page 44

448-2 Calibration Procedure (For all models) Note As to the correction value used for one-point calibration and two-point calibration, enter the pr

Page 45

458-2-1 One-point Calibration 1 In the standby mode, press and hold the HOLD key to enter the calibration mode. " Cal " appears. 2 Sele

Page 46

468-2-2 Two-point Calibration 1 In the standby mode, press and hold the HOLD key to enter the calibration mode. " Cal " appears. 2 Sele

Page 47 - 8-2-1 One-point Calibration

47 9 When the measurement of the first point has completed, clean the sensor plates, temperature sensor and protector and prepare the second standar

Page 48 - 8-2-2 Two-point Calibration

31. INTRODUCTION This manual describes how the SV-A/SV-H series viscometer works and how to get the most out of it in terms of performance. Read th

Page 49

488-2-3 Simplified Calibration Using Purified Water (SV-1A/1H/10A/10H) 1 Place the purified water in the sample cup. 2 Press the START key to m

Page 50 - of the purified

49 Reference data: Theoretical viscosity value (Viscosity × Density) of the purified water at various temperatures Temperature (°C) Viscosity × D

Page 51

509. FUNCTION SETTING (FOR ALL MODELS) The viscometer, by selecting functions to be used in the function setting, can specify the performance approp

Page 52

51Example of the function setting procedure To change the setting of "Unit upon power-on (Unit)" to the viscosity: cP (Centipoise) and the

Page 53

529-2 Details of the Function Items (For all models) Function item Parameter Description 5vtyp Htype • Portable type(This function can use: H-fnc

Page 54

53S5t-b 0 Stable when within±5% 1 • Stability band width 2 Stable when within±15% When setting the auto hold function, it can be set.(H-fnc 1) Hld

Page 55

549-3 Description of Items (For all models) Condition (Cond) The stability of the viscosity measurement results can be adjusted, taking ambient cond

Page 56

55MODE Key Function During Measurement (fnc) Parameter Settings Description 0 • Switches the viscosity units. Each time the MODE key is pressed, the

Page 57

56Data Output Format (type) The output format appropriate for the device connected to RS-232C can be selected. Parameter Settings Description 0 A&

Page 58

57Other Output (5-ed) Parameter Settings Description 0 No output 1 Outputs remarks. 2 • Outputs remarks, Device ID information and signature. With

Page 60 - Setting example

58Comparator (Cp) The results of the comparison are indicated by the sound. (It is necessary to set : bep- , bep-, bep-) To compare, use:  Upper

Page 61 - Press and hold

59Setting the ID number 1 In the standby mode, press and hold the MODE key to enter the function setting mode. " 5vtyp " appears. 2 Pre

Page 62

60Initialization (Clr) Restores the following data to the default setting.  Function setting (Setting of “5vtyp” is not returned by this operation)

Page 63 - Date/Time (Cladj)

61Date/Time (Cladj)  The upper two digits of the year are not displayed. For example, the year 2007 is displayed as "07".  The time is se

Page 64

62Changing the date The date is changed in the selected displaying order. The following is an example when the order of " y " (Year), "

Page 65 - (Go to page 61)

63Changing the time 13 The current time is displayed. Press the MODE key to display the current date. 14 Press the PRINT key to select the settin

Page 66

649-4 Data Output Format Examples (For all models) 9-4-1 A&D Standard Format Used with the printer MODE 1 or MODE 2 when the optional compact

Page 67

65SV-10A/SV-10H output format example Viscosity unit Display Output format Remarks LLLLL mPa・s OL,-99999999mPs Below measuring range error 0.30m

Page 68 - 9-4-2 D.P. Format

66SV-100A/SV-100H output format example Viscosity unit Display Output format Remarks L Pa・s OL,-99999999Pas Below measuring range error 1.00

Page 69 - 9-4-3 CSV Format

67Printing format example (3) Function setting (√=Output Blank=No output) 5-at 1 Measurement elapsed time √ 5-td 1 Date/time √ Remarks Device ID i

Page 70

51-2 Features  High accuracy The Vibro Viscometer adopting the sine-wave vibration technique (SV type), achieves a high measurement accuracy of 1%*1

Page 71

68Output format example (1) With ID number, date and time added Function setting (√=Output Blank=No output) 5-td 1 Date/time √ 5-ed 3 Device ID nu

Page 72

69 SV-10A/SV-10H output format example Viscosity / Temper-ature Display Output format example Remarks LLLLLmPa・s LAB-12,2003/03/19,12:34:56,+025.67

Page 73 - ,,,+025.67,C,+00001.23, Pa s

70SV-100A/SV-100H output format example Viscosity / Temper-ature Display Output format example Remarks L Pa・s LAB-12,2003/03/19,12:34:56,+025

Page 74

71SV-1A/SV-1H/SV-10A/SV-10H output format example Viscosity / Temperature Display Output format example mmPa⋅s / °C 1.23 mPa⋅s ,,,+025.67,C,+0000

Page 75

72SV-1A/SV-1H output format example Viscosity / Temper-ature Display Output format example Remarks LLLLL mPa・s +00000.00,mPa s,+025.67,C Zeroes ar

Page 76 - 10-1 Introduction

73SV-10A/SV-10H output format example Viscosity / Temper-ature Display Output format example Remarks LLLLL mPa・s +00000.00,mPa s,+025.67,C Zeroes

Page 77

7410.CONNECTION TO A PERSONAL COMPUTER 10-1 Introduction If connecting the SV-A / SV-H series to a personal computer using the Windows communicatio

Page 78 - Tab, and click

7510-3 Connection to a personal computer In a case where the personal computer has a COM port The SV-A / SV-H seri

Page 79

7610-4 Configuration of the COM port When connecting the SV-A / SV-H series to a personal computer, confirm that the COM port setting match up. (

Page 80

7710-5 Controlling the measurement using a personal computer (In case of the graphing software“RsVisco”

Page 81 - Display unit rear side

6  Calibration The viscometer can be calibrated using a standard viscosity fluid or a sample of a known viscosity. Calibration allows the viscomete

Page 82 - (FOR ALL MODELS)

7811. CONNECTION TO A PRINTER (FOR ALL MODELS)  The viscometer can be connected to the optional compact printer AD-8121B using the standard RS-232C

Page 83 - 14. TROUBLESHOOTING

7912. RS-232C SERIAL INTERFACE (FOR ALL MODELS) RS-232C Serial Interface Transmission system EIA RS-232C Transmission form Asynchronous, bi-directi

Page 84

8013. COMMAND LIST (FOR ALL MODELS) The viscometer can be controlled by the following commands from the computer. Add a terminator ( 0Dh、0Ah ) to ea

Page 85 - (For all models)

81ProtectorRotationalaxisSensor unit121Surfacelocator plateSensor protectivecoverTemperature sensor14. TROUBLESHOOTING (FOR ALL MODELS) The viscomete

Page 86

82 (2) How to remove and attach the surface locator plate (Only for SV-10A/SV-10H/SV-100A/SV-100H) Removing Loosen the screw and remove the surface

Page 87 - 15. ERROR DISPLAY

8314-2 When measurement values are not correct (For all models)  Is the viscometer in the data hold mode?  To release the data hold mode, press t

Page 88 - 16. SPECIFICATIONS

84 Sample cup influence The viscometers have been calibrated with the following cups when shipped. When using another cup, use that cup to measur

Page 89

8515. ERROR DISPLAY (FOR ALL MODELS) Error display Description Above measuring range error The viscosity value exceeds the upper limit of the visc

Page 90 - 17. OPTIONAL ACCESSORIES

8616. SPECIFICATIONS Measurement method SV-1A/SV-1H SV-10A/SV-10H SV-100A/SV-100H Measurement method Sine-wave Vibro Viscometer using the Tuning-for

Page 91 - AX-SV-33 Sample Cup

87 *1 For the SV-1A/1H when a sample cup of 2 mL is used. For the SV-10A/10H and SV-100A/100H when a sample cup of 45 mL is used. *2 Repetitive m

Page 92

72. UNPACKING THE VISCOMETER  The viscometer is a precision instrument. Unpack the viscometer carefully. Keep the packing material to be used for t

Page 93

8817. OPTIONAL ACCESSORIES List of Optional Accessories (sold separately) Available () Model Name 1A/1H 10A/10H 100A/100HAX-SV-31-2.5 Standard vis

Page 94 - Plastic storage container

89Available () Model Name 1A/1H 10A/10H 100A/100HCup set(Capacity: 2 mL) (Water jacket attached) ( Used with the all model of the SV series) AX-SV-5

Page 95 - Serial-USB converter

90 AX-SV-36 Positioning Stopper • Used to fix the position of the sensor unit and the sensor plates so that positioning the sensor plates and the sa

Page 96

91 AX-SV-37 Water Jacket Assembly • Used, in combination with a commercially available constant temperature bath for heating medium circulation, to

Page 97 - Sample cup

92 AX-SV-38 Glass Storage Container • Used to store sample fluids. The viscosity of the sample stored in the container can be measured as it is.

Page 98

93 AX-SV-52 X-Y-Z Stage • Used to support and adjust the sample cup position. - Consists of : X-Y-Z stage 1 pc AX-SV-53-EX Software Set •

Page 99 - (Lid attached)

94 AX-SV-54 Cup Set (10 mL / 13 mL / 45 mL) (Use the AX-SV-51 Stand Set if necessary.) • Used when a small amount of sample will be measured. - C

Page 100 - Sensor unit (Handle*)

95Only for SV-1A/SV-1H (When using this option, use the AX-SV-51 Stand Set) AX-SV-55 Cup Set (2 mL) • Sample cup set (Only for SV-1A/SV-1H) Used

Page 101

96 AX-SV-56-1 Sample Cup Holder (For 2 mL capacity, Transparent) AX-SV-56-2 Sample Cup Holder (For 2 mL capacity, Black) • Used for securing a

Page 102

97 AX-SV-60 Square glass sample Cup (10 mm x 10 mm x 45 mm) • Can be used for optical measuring devices such as a spectrophotometer. - Consis