A&D HT-120 Specifications Page 9

  • Download
  • Add to my manuals
  • Print
  • Page
    / 12
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 8
9
CALIBRATION
When is Calibration Required?
Calibration may be required when
the scale was initially installed or
has been moved to another
location.
Calibration Using a Calibration Weight
Step 1 Press the Calibration switch while the scale is in the weighing mode.
Then, Cal will be displayed.
Step 2 Press the RE-ZERO key and Cal 0
g will be displayed.
With nothing on the weighing pan, press the RE-ZERO key.
Step 3 The display shows Cal blinking while waiting for the weighing
value to be stable. When becoming stable, the scale perform zero
calibration and will show Cal100
g (factory default) with tens place
digit blinking).
Select the next step as follows:
Proceed to the span calibration of Step 4.
Press the ON:OFF key to finish the procedure without
performing the span calibration.
Step 4 At the display Cal100
g , place the calibration weight 100 g (±2 mg)
on the center of the weighing pan and press the RE-ZERO key.
To use the other calibration weight (50 g or more recommended),
press the UNITS key to change by 10 g and place the weight
equal to the displayed value.
Step 5 The display shows Cal blinking while waiting for the weighing
value to be stable. When becoming stable, the scale perform span
calibration and will show end . Remove the calibration weight
and press the ON:OFF key to finish the procedure.
Gravity Acceleration Compensation
The scale is calibrated with the gravity acceleration value 9.7985 m/s
2
at the factory. When the gravity acceleration at your location is
different value, you should calibrate the scale using a calibration
weight. But if the calibration weight cannot be prepared, change the
Calibration
switch cover
Calibration switch
Page view 8
1 2 3 4 5 6 7 8 9 10 11 12

Comments to this Manuals

No comments