7. Check Mode
7.1. Checking the Modbus-RTU Communication Status
7.1.1. Entering the Check Mode
Step 1 While pressing and holding the ENTER key, press the F key.
fnc is displayed to indicate that the indicator will enter the general function
mode.
To go back to the weighing mode, press the ESC key.
Step 2 While pressing and holding the ZERO key, press the ENTER key.
1Chc is displayed to indicate that the indicator will enter the check mode.
Press the ENTER key again to display an item to be checked.
Step 3 Press the or key to select Chcrtu (Modbus check mode) and press the
ENTER key to enter the Modbus check mode.
To exit from the check mode, press the ESC key.
Table 12 Check mode list
Display Checking item
Chckey
Key switches
Chc Cl
Standard serial output
Chc∗∗∗
Chcrtu
Chc∗∗∗
Interfaces
Modbus-RTU
Chc r5
Testing terminal
Chc ad
A/D (Load cell)
Chc in
Internal count
Chcpr9
Program version
Chc 5n
Serial number
C5 pr9
Program checksum
C5 fra
Memory (FRAM) checksum
Calfdt
Calibration-related functions